Getting Started Model 9110EH Instruction Manual
Set the expected NO and NOx span gas concentration. If you supply NO gas to the
analyzer, the values for expected NO and NO
x
need to be identical.
This sequence causes the
analyzer to prompt for the
expected NO and NO
x
span concentration.
EXIT ignores the new setting and
returns to the CONCENTRATION
MENU.
ENTR accepts the new setting and
returns to the
CONCENTRATION MENU..
M-P CAL CONCENTRATION MENU
NOX NO CONV
EXIT
M-P CAL NOX SPAN CONC: 400.0 Conc
0 0 0 4 0 .0 ENTR EXIT
M-P CAL CONCENTRATION MENU
NOX NO CONV
EXIT
M-P CAL NO SPAN CONC: 400.0 Conc
0 0 0 4 0 .0 ENTR EXIT
M-P CAL RANGE = 500.0 PPB NO = X.X
< TST TST > ZERO CONC
EXIT
SAMPLE RANGE = 500.0 PPB NOX= X.X
< TST TST > CAL SETUP
The NO
x
& NO span concentration
values automatically default to
400.0 Conc.
To change this value to the actual
concentration of the span gas, enter
the number by pressing the key
under each digit until the expected
value appears.
28 M9110EH Rev 0