Ain calibration, Test channel output, Table 6-12 – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual
Page 86: Current loop output calibration with resistor

Operating Instructions Model 9110EH Instruction Manual
Table 0-12: Current Loop Output Calibration with Resistor
Full scale
Voltage for 2-20 mA
(measured across resistor)
Voltage for 4-20 mA
(measured across resistor)
0%
0.5 V
1.0 V
100%
5.0 V
5.0 V
6.7.3.6. AIN Calibration
This is the sub-menu to conduct the analog input calibration. This calibration should only be
necessary after major repair such as a replacement of CPU, motherboard or power supplies.
Exit at any time to
return to the main
DIAG menu
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
DIAG AIO
AIN A/C FREQUENCY: 60 HZ
SET> EDIT
EXIT
DIAG AIO
AIN CALIBRATED: NO
< SET SET> CAL
EXIT
DIAG AIO
AIN CALIBRATED: YES
< SET SET> CAL
EXIT
Instrument
calibrates
automatically
Exit to return to the
ANALOG I/O
CONFIGURATION
MENU
STARTING FROM ANALOG I / O CONFIGURATION MENU
DIAG AIO
CALIBRATING A/D ZERO
CALIBRATING A/D SPAN
6.7.4. Test Channel Output
When activated, output channel A4 can be used to report one of the test functions viewable
from the SAMPLE mode display.
72 M9110EH Rev 0