Teledyne GC-Pro/FID - Gas Chromatograph User Manual

Page 55

Advertising
background image

GC-Pro FID

Operation

Teledyne

Analytical

Instruments

41

4.7.1 Default Parameters

The versatility of this analyzer usually results in significant changes

being made to parameters over the course of time to better suit a
particular application. Occasionally processes change requiring
alteration to alarms, filter settings etc. At some time, it may be
beneficial to reset the analyzer to the default conditions as it was when
shipped from the factory. Below is a listing of the default parameters
used in configuring the typical GC-Pro FID instrument. Your specific
application may require different settings. Refer to the Addendum and
Testing Results
section of Appendix C for any specific changes or
recommendations that apply to your application.

Range/Application:

Refer to the Addendum and Testing
Results
section of Appendix C

Range: Manual

Alarms:

Defeated, 20ppb, 50ppb, HI, NON-
FAILSAFE, NON-LATCHING

Auto Span Timing:

Defeated, every 7 days, at 12 hours

Span: 80

ppb

Password: 11111

4.7.2 The HOME Screen

The HOME screen is the default display and represents the Analysis

Mode of the instrument. It is a set of two screens that present
information regarding the current measurement state. The HOME screen
is where the maximum information about the instrument is displayed.

The Group name/tag is listed on the left most column and the

concentration for groups are displayed in PPB, PPM or % depending on
the range and value. The Enabled group is represented by the glowing
LED. Each group has three user settable ranges R1, R2, and R3 and the
current range index is displayed in the Group row. It is possible to view
the VS (Volt-second) count by touching the “Vs Count” button.

Advertising