Any group that is available (see Section 4.7.8 Group Setup) can be
assigned to any output channel.
To assign a group to a channel: Press the UP Arrow button under
group and it will start with G00, pressing UP again increments it to G01,
and so on. After G09, it will return to G00.
4.7.10 Calibration
The SPAN function is used to calibrate the analyzer. Span can be
performed either manually or automatically with the auto calibration
function if equipped. Typically, each group will require a separate
calibration.
The analyzer is calibrated using span gas as described in Section 4.1.
See also Appendix C Addendum and Testing Results for any specific
requirements regarding span gas for your application. This section
assumes that this gas has been properly connected and the line checked
for leaks.
To initiate a span calibration, touch the Calibration icon on the
MAIN screen. The following screen is displayed.