Operation
Teledyne
Analytical
Instruments
51
After modifying the above parameters touch the ENTER button to
save the new inputs. A confirmation window will pop up before actually
saving the inputs. Select OK to save, otherwise touch CANCEL to reject
the new inputs.
4.7.7 Temperature
Temperature control is a critical factor in GC analysis. The GC-Pro
FID Analyzer incorporates three dedicated temperature controllers.
FID Temperature (Sensor)
GC Column Temperature (Column)
Sample Chamber (Sample)
Refer to the Addendum and Testing Results section of Appendix C
for the specific temperature setpoints used for your system and
application. Do not change these values unless directed by Customer
Service at TAI. For further information please contact TAI Customer
Service.
Caution:
The temperature settings have been carefully
determined and set at the factory for your
application. Temperature is a key factor in
determining the accuracy of the resultant
analysis. It is recommended that the user consult
TAI before making any changes to the
temperature setpoints as listed in Appendix C of
this manual.