Teledyne GC-Pro/FID - Gas Chromatograph User Manual

Page 95

Advertising
background image

GC-Pro FID

Operation

Teledyne

Analytical

Instruments

81

This screen displays the concentration behavior or trend for the

group on y axis against the time into the cycle on the x axis.

Low and High values can be set for the y axis using the “L” and “H”

boxes below the graph. Touching the “Update” button will reformat the
y axis index with the new input.

It is possible to view the instantaneous value by touching the trend

graph at the location of interest and the corresponding value will be
displayed in “Watch Line” box at the top right corner.

Note: The LOW and HIGH values must be updated by the user

as soon as system is powered on.

Note: The trend graph will not be plotted continuously; it needs to

be refreshed manually. To refresh the trend screen, press
the HOME button to return to the HOME screen and then
re-enter the trend screen by pressing trend button.

4.7.13.2

T_C

YCLE

Similar to the above function, the T_Cycle button displays a graph of

group concentration versus cycles.

The T_Cycle is display is shown below.


Both X-axis (number of cycles) and y-axis (concentration) can be

changed by touching the “Setting” button. The maximum number of
cycles that can be plotted is 300.

Advertising