6.10.9.4. Changing the Analyzer’s HOSTNAME
The HOSTNAME is the name by which the analyzer appears on your network. The default name
for all Teledyne Instruments Model GFC7000E analyzers is MGFC7000E. To change this name
(particularly if you have more than one Model GFC7000E analyzer on your network), press.
SETUP X.X
COMMUNICATIONS MENU
ID
INET
COM1
EXIT
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL
SETUP
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS DIAG ALRM
EXIT
SETUP X.X
DHCP: ON
SET>
EDIT
EXIT
SETUP X.X
HOSTNAME: M360E
<SET
EDIT
EXIT
SETUP X.X
HOSTNAME: [M]360E
<CH CH> INS DEL [?]
ENTR EXIT
SETUP X.X
HOSTNAME: M360E-FIELD1
<SET EDIT
EXIT
Continue pressing SET> UNTIL …
SETUP X.X
INITIALIZING INET 0%
…
INITIALIZING INET 100%
SETUP X.X
INITIALIZATI0N SUCCEEDED
SETUP X.X
INITIALIZATION FAILED
SETUP X.X
COMMUNICATIONS MENU
ID
INET
COM1
EXIT
Contact your IT
Network Administrator
Table 6-19: Internet Configuration Keypad Functions
KEY FUNCTION
<CH
Moves the cursor one character to the left.
CH>
Moves the cursor one character to the right.
INS
Inserts a character before the cursor location.
DEL
Deletes a character at the cursor location.
[?]
Press this key to cycle through the range of numerals and characters available
for insertion.
0-9, A-Z, space ’ ~ ! © # $ % ^ & * ( ) - _ = +[ ] { } <
>\ | ; : , . / ?
ENTR
Accepts the new setting and returns to the previous menu.
EXIT
Ignores the new setting and returns to the previous menu.