Sample chamber, Figure 3-12: optical bench layout – Teledyne GFC-7000E - Trace CO2 Analyzer User Manual
Page 42

Model GFC7000E Instruction Manual
Getting Started
04584 Rev A1
31
GFC Temperature
Sensor
Shock Absorbing
Mounting Bracket
Purge Gas
Inlet
IR Source
GFC Wheel
Heat Sync
GFC Wheel Motor
Purge Gas
Pressure Regulator
Bench
Temperature
Thermistor
Sample Chamber
Pressure Sensor(s)
Sample Gas Outlet
fitting
Sample Gas Flow
Sensor
Sync/Demod PCA
Housing
Opto-Pickup
PCA
GFC Heater
Figure 3-12: Optical Bench Layout
GFC Wheel
Motor
PARTICULATE
FILTER
INSTRUMENT CHASSIS
PUMP
GFC Wheel
Housing
SAMPLE CHAMBER
GFC Motor
Heat Sync
VENT SPAN
OUTLET
PRESSURE
SPAN INLET
IZS INLET
EXHAUST GAS
OUTLET
PURGE GAS
INLET
SAMPLE GAS
INLET
Purge Gas
Flow Rate
Control
Orifice
Purge Gas
Pressure
Control Assy
S
a
m
p
le
G
a
s
C
rit
ic
a
l
F
lo
w
O
ri
fice
FLOW
SENSOR
SAMPLE
PRESSURE
SENSOR
FLOW / PRESSURE
SENSOR PCA
Figure 3-13: Internal Pneumatic Flow – Basic Configuration