Maintenance – General Technologies CT8017 Autorange Digital Multimeter User Manual

Page 2

Advertising
background image

4.3.1 AC/DC Voltage measurement

Connect the black test lead to the “COM” socket and red test lead
to the “V

mA” socket.

Set the FUNCTION SWITCH to the desired range:

V” function for Manual range

V” function for Auto range

If MANUAL MODE is selected and the voltage range is not

known beforehand, using the RANGE BUTTON to set the range
to the highest and work down as needed.

If in MANUAL MODE, select AC or DC voltage measurement by
pressing the AC/DC/Beeper/Diode Button function button.

Connect the test leads across the source or load under
measurement.

The measurement and polarity will be shown the LCD Display
when the probe are connected.

4.3.2 AC/DC Current measurement

Connect the black test lead to the “COM” socket and red test lead to:

V

mA” for measurements below 320 mA

10 A” for measurements between 320 mA and 10 A

Select the current range using the FUNCTION SWITCH, if the

current range is not known beforehand, start with the highest
range and work down as needed.

If needed select AC or DC current measurement by pressing the

AC/DC/Beeper/Diode Button function button.

Connect the test leads in series with the circuit or load under
measurement.

The measurement and polarity will be shown the LCD Display
when the probe are connected and current flows through the
meter.

4.3.3 Resistance measurement

CAUTION: Maximum Input Voltage for this function is 250

Vrms for less than 10 Sec., do not exceed this
rating to avoid personal injuries or damage to the
instrument. Also ensure there is no power applied
to the component or circuit and all capacitors are
discharged.

Connect the black test lead to the “

” socket and red test

lead to the “V

mA” socket.

Set the FUNCTION SWITCH switch to the “

Ω” function.

Connect the test leads across the componet or circuit under
measurement.

• The range will adjust automatically for optimal readout,

and

the measurement will be shown the LCD Display when the

probe are connected.

4.4 Other Functions

CAUTION: Maximum Input Voltage for this function is 250

Vrms for less than 10 Sec., do not exceed this
rating to avoid personal injuries or damage to the
instrument. Also ensure there is no power applied
to the diode.

4.4.1 Diode test

Connect the black test lead to the “COM” socket and red test lead
to the “V

mA” socket.

Set the FUNCTION/RANGE switch to the “

” function.

Press the AC/DC/Beeper/Diode button until the diode test

enable symbol

appears on the display

Proceed to connect the test leads across the diode observing the
polarity: red probe to the anode (+) of the diode and black test
lead to the cathode (-).

The measurement will be shown on the LCD Display when the
diode is connected.

4.4.2 Continuity Test

Connect the black test lead to the “COM” socket and red test lead
to the “V

mA” socket.

Set the FUNCTION/RANGE switch to the “

” function.

Press the AC/DC/Beeper/Diode button until the beeper enable

symbol appears on the display

Proceed to connect the test leads across the circuit or component to test.

Buzzer will sound if the circuit resistance is below 18

4.4.3 Display Hold

The Hold Button is used to hold display readings during

measurements.
• Pressing the button holds the display reading, and “H” appears

on the display.

• Pressing the button again resumes normal operation.

5. MAINTENANCE

CAUTION: Before attempting battery removal or replacement,

disconnect test leads and remove the instrument
from any energized circuit to avoid shock hazard.

5.1 Battery Replacement

To replace the battery, remove the screw of the back Battery Cover
(back of the case) and remove the batteries.

Replace with new 1.5 V alkaline battery type AA, UM3 or equivalent
observing the proper polarity from the diagram on the label inside
the battery compartment.

Reinstall the battery cover and tighten the securing screw.

5.2 Fuse replacement

Remove the screw of the back Battery Cover (back of the case)
and remove the batteries.

Remove the four screws at the bottom of the case and of the bat-
tery compartment.

Replace new fuses only with the identical type and rating.

F1=200mA: Type IEC60127-2 or UL248-14 (5 x 20mm)

fast acting fuse, rated at 400mA/250V.

F2=10A: Type IEC60127-2 orUL248-14 (6.3x25.4mm)

fast acting fuse, rated at 10A/250V.

Reinstall the back cover, and replace the four screws.

Reinstall the battery observing the polarity on the battery compartment
label, the battery cover and tighten the securing screw.

5.3 Cleaning

Periodically wipe the case with a soft damp cloth and mild household
cleanser. Do not use abrasives or solvents. Ensure that no water gets
inside the instrument to prevent possible shorts and damage.

6. WARRANTY

One year limited warranty, excluding batteries and fuses. For details
see Standard Warranty Information in our webpage or you may request
a printed copy.

© Copyright 2001 General Technologies

General Technologies Corp.

#121 - 7350 72nd Street Tel.: (604) 952-6699

Delta, BC

Fax: (604) 952-6690

Canada V4G 1H9

www.generaltechnologies.net

Advertising