Maintenance – General Technologies CT8022 Ergonomic Autorange Digital Multimeter User Manual

Page 2

Advertising
background image

4.3.1 DC Voltage measurement

Set the FUNCTION SWITCH to “V” and the RANGE SWITCH to
DC/

function.

Connect the test leads across the source or load under
measurement.

The measurement and polarity will be shown the LCD Display
when the probes are connected.

4.3.2 AC Voltage measurement

Set the FUNCTION SWITCH to “V” and the RANGE SWITCH to
AC/ ” function.

Connect the test leads across the source or load under
measurement.

The measurement will be shown on the LCD Display when the
probes are connected and current flows through the meter.

4.3.3 DC Current measurement

Set the FUNCTION SWITCH to “A” and the RANGE SWITCH to
DC/

function.

Connect the test leads in series with the source or circuit under
measurement.

The measurement and polarity will be shown the LCD Display
when the probes are connected and current flows through the
meter.

4.3.4 AC Current measurement

Set the FUNCTION SWITCH to “A” and the RANGE SWITCH to
AC/ ” function.

Connect the test leads in series with the source or load under
measurement.

The measurement will be shown on the LCD Display when the
probes are connected and current flows through the meter.

4.3.5 Resistance measurement

CAUTION: Maximum Input Voltage for this function is 250

Vrms for less than 10 Sec., do not exceed this
rating to avoid personal injuries or damage to the
instrument. Also ensure there is no power applied
to the component or circuit and all capacitors are
discharged.

Set the FUNCTION SWITCH to “R” and the RANGE SWITCH to
DC/

function.

Connect the test leads across the component or circuit under
measurement.

The range will adjust automatically for optimal readout, and the

measurement will be shown the LCD Display when the probe are

connected.

4.4 Other Functions

CAUTION: Maximum Input Voltage for this function is 250

Vrms for less than 10 Sec., do not exceed this
rating to avoid personal injuries or damage to the
instrument. Also ensure there is no power applied
to the diode.

4.4.1 Continuity Test

Set the FUNCTION SWITCH to “R” and the RANGE SWITCH to
AC/ ” function.

Connect the test leads across the circuit or component to test.

Buzzer will sound if the circuit resistance is below 100Ω.

4.4.2 Logic Level Test

Set the FUNCTION SWITCH to “TTL” and the RANGE SWITCH
to “DC/

function.

Connect the test leads across the source or circuit under test.

The logic level will be shown as “HI” or “LO” on the LCD display
according to the table below:

© Copyright 2001 General Technologies

General Technologies Corp.

#121 - 7350 72nd Street Tel.: (604) 952-6699

Delta, BC

Fax: (604) 952-6690

Canada V4G 1H9

www.generaltechnologies.net

* The display will switch between HI and LO, but the response is

affected by the pulses frequency

4.4.3 Display Hold

A push button switch at the back of the unit is used to hold display

readings during measurements.
• Pressing the button holds the display reading, and “DH” appears

on the display.

• Pressing the button again resumes normal operation.
4.4.4 Auto Power OFF
The meter will turn off automatically to extend the battery life after

approx. 30 minutes without operating. To turn it on again, set

the FUNCTION SWITCH to the “OFF” position and then to the

desired range position.

5. MAINTENANCE

CAUTION: Before attempting battery removal or replacement,

disconnect test leads and remove the instrument
from any energized circuit to avoid shock hazard.

5.1 Battery Replacement

• When the battery needs replacement, the battery symbol

will

appear in the lower left hand side of the LCD display.

• To replace the battery, remove the screw at the back Battery

Cover and replace with a new 9 V alkaline battery type NEDA
1604 or IEC6F22, observing the polarity as indicated at the back
of the case.

• Reinstall the battery cover and tighten the securing screw.

5.2 Fuse replacement

• To replace the internal fuse, turn the Fuse Cap counter clockwise,

to release the Fuse Cap assembly.

• Take out the blown fuse and replace with a type IEC60127-2 or

UL248-14 (5 x 20mm) fast acting fuse, rated at 400mA/250V.

• Replace the fuse cap, by inserting it in the meter body and turning

the Fuse Cap assembly clockwise, unitl it locks.

5.3 Cleaning

Periodically wipe the case with a soft damp cloth and mild household
cleanser. Do not use abrasives or solvents. Ensure that no water gets
inside the instrument to prevent possible shorts and damage.

6. WARRANTY

One year limited warranty, excluding batteries and fuses. For details
see Standard Warranty Information in our webpage or you may request
a printed copy.

Input Voltage

TTL Logic Status

Display

Above 2.0 V

High

HI

Below 0.8 V

Low

LO

0.8 to 2.0 V

Undertermined

-

Negative input (-V)

Undertermined

HI/LO or LO/HI*

Pulse

Undertermined

HI/LO or LO/HI*

Advertising