Gentec-EO UM-B User Manual

Page 12

Advertising
background image

12

Revision 2.1

UM-B Series Instruction Manual

Gentec Electro-Optics Inc. All rights reserved

Gaussian beam stopped at 1/

e

2

) applied to a diameter equal to 80% of the

detector aperture. The use of a divergent lens, a Lambertian diffuser such as opal
glass, or any other method of beam spreading, is recommended for this purpose.
Please take note that all of the laser light must be directed within the detector
aperture and that the transmission loss through the optical component must be
known.



6- Adjust the oscilloscope so it triggers on the wattmeter pulse or on the chopper sync. signal.


7- Measure the baseline to peak voltage generated by the wattmeter.


8- Determine the wattmeter Volt/watt sensitivity from the detector identification label or calibration

certificate. Choose the value stated for the wavelength being used.

9- Calculate the optical energy using the following equation:


Power = (V

peak

-V

offset

) / Calibration sensitivity

Ex:

- V

peak

-V

offset

= 1 volt

- Detector calibration sensitivity (100 Volts / watt)


Energy = 1 Volt / 100 V/W = 10 mW

NOTE: Exclude any DC offset from the pulse peak value measurement.























Advertising