3.3.2 Working at other wavelengths than 1.064µm

For measurements with a UM-B Series

wattmeter at wavelengths other than 1.064µm, a

correction factor must be set to compensate for the change in sensitivity of the wattmeter
caused by the change in absorption of the optical absorber at different wavelengths.

To correct for the change in absorption refer to the spectral curve

of the “ Personal

Wavelength Correction

TM

“ certificate supplied for the wattmeter and calculate K by taking

the percentage difference between the absorption @1.064µm and that at the desired
wavelength.

 

μm

064

.

1

@

1

A

A

K

Power = (V

peak

-V

offset

) / Calibration sensitivity / K

Here

 

1

A

=

Absorption of the UM-B

@ the desired wavelength.

μm

064

.

1

@

A

=

Absorption of the UM-B

@ 1.064µm

A sample calculation follows:

 

1

A

=

92 %.

μm

064

.

1

@

A

=

94 %

K =

 

μm

064

.

1

@

1

A

A

x 100

K =

%

94

%

92

x 100 = 0.9787 x 100 = 97.87 %


Ex:

-

(V

peak

-V

offset

) = 1 volt

-

Detector calibration sensitivity @1.064µm (100 Volts / Watt)


Power = 1 Volt / 100 V/W / 97.87% = 10.218 mW