Agilent 8753es specifications, Test-port output characteristics, Test-port input characteristics – Agilent Technologies 8753ET User Manual

Page 10

Advertising
background image

10

Test-Port Output Characteristics

7

Frequency characteristics
Range

30 kHz to 3 GHz

Option 006

30 kHz to 6 GHz

Resolution

1 Hz

Stability

9

±7.5 ppm (0° to 55° C)
±3 ppm/year

Option 1D5

±0.05 ppm (0° to 55° C),
±0.5 ppm/year

Accuracy

±10 ppm at 25° ±5° C

Power range

2

-85 to +10 dBm

Option 075/014

-85 to +8 dBm

Power sweep range

25 dB (typically 31 dB)

Resolution

0.01 dB

Level accuracy

1,2,5

±1.0 dB

Level linearity

1,2,5

-15 dBm to +5 dBm ±0.2 dB
5 dBm to 10 dBm

6

±0.5 dB

Impedance

9

50

Ω

(nominal)

30 kHz to 3 GHz

>16 dB RL (<1.38 SWR)

3 GHz to 6 GHz

>14 dB RL (<1.50 SWR)

Option 075

75

Ω

(nominal)

30 kHz to 3 GHz

>16 dB RL (<1.38 SWR)

Spectral purity
2nd harmonic

3

<-40 dBc at 0 dBm

9

<-50 dBc at -10 dBm

9

Option 002

<-25 dBc at 10 dBm

6

3rd harmonic

4

<-40 dBc at 0 dBm

9

<-50 dBc at -10 dBm

9

Option 002

<-25 dBc at 10 dBm

6

Nonharmonic spurious

9

Mixer related

<-30 dBc at 10 dBm

6

<-55 dBc at -10 dBm

Test-Port Input Characteristics

Frequency range

30 kHz to 3 GHz

Option 006

30 kHz to 6 GHz

Average noise level

2,8

<3 GHz

<-82 dBm (3 kHz BW)

<3 GHz

<-102 dBm (10 Hz BW)

<3 GHz

<-110 dBm (10 Hz BW) typical

3 to 6 GHz

<-77 dBm (3 kHz BW)

3 to 6 GHz

<-97 dBm (10 Hz BW)

3 to 6 GHz

<-105 dBm (10 Hz BW) typical

Maximum input level

10 dBm

Damage level

26 dBm or 35 VDC

Impedance

50

Ω

(nominal)

Option 075

75

Ω

(nominal)

Frequency response

2,5

300 kHz to 3 GHz

±1.0 dB

3 GHz to 6 GHz

±2.0 dB

Internally generated harmonics (Option 002)
2nd harmonic

3

<-15 dBc at +8 dBm
<-30 dBc at 0 dBm

9

<-45 dBc at -15 dBm

9

3rd harmonic

4

<-30 dBc at +8 dBm
<-50 dBc at 0 dBm

9

<-50 dBc at -15 dBm

9

Harmonic measurement accuracy

10

300 kHz to 3 GHz

±1.5 dB

3 GHz to 6 GHz

±3 dB (Option 006)

Harmonic measurement dynamic range
-40 dBc (output = -10 dBm, input = <-15 dBm)

9

Frequency offset mode

11

Frequency range

300 kHz to 3 GHz

Option 006

300 kHz to 6 GHz

R-channel input requirements

Power level
300 kHz to 3 GHz

0 to -35 dBm

3 GHz to 6 GHz

0 to -30 dBm

LO spectral purity (typical)

Maximum spurious input <-25 dBc
Residual FM

<20 kHz

LO frequency accuracy

9

±1 MHz at nominal frequency

Agilent 8753ES Specifications

1. Relative to 0 dBm output power.
2. Typical below 300 kHz.
3. 16 MHz to 3 GHz.
4. 16 MHz to 2 GHz.
5. Typical from 2 to 3 GHz for instruments with Option 075.
6. +8 dBm maximum with Option 075, or Option 014.
7. Test performed on port 1 only.
8. Instruments with Option 075 are degraded 2 dB.
9. Typical performance.
10. 25° ±5° C.
11. The Agilent 8753ES source characteristics and measurement accuracy in this

mode are dependent on the stability of the external LO source. The RF source
tracks the LO to maintain a stable IF signal at the R-channel receiver input.
Degradation in accuracy is negligible when using an Agilent 8642A/B, 8656B,
or E4432B RF signal generator as the LO source.

Advertising