Magnitude characteristics, Phase characteristics – Agilent Technologies 8753ET User Manual
Page 15

15
External source mode
1
(CW time sweep only)
Frequency range
300 kHz to 6 GHz
R-channel input requirements
2
Power level
0 to -25 dBm
Spectral purity
Maximum spurious input <-30 dBc
Residual FM
<20 kHz
Typical settling time
500 ms (automatic)
50 ms (manual)
Frequency readout accuracy
2
0.1% (automatic)
Input frequency accuracy requirement
2
Manual
-0.5 to 5 MHz
Accuracy
(See magnitude and phase
characteristics)
Magnitude characteristics
Dynamic accuracy
(10 Hz IF BW)
Display resolution
0.001 dB/division
Marker resolution
3
0.001 dB
Trace noise
(0 dBm at transmission port or +5 dBm at reflection port,
ratio measurement, 3 kHz BW)
300 kHz to 3 GHz
<0.006 dB rms
3 GHz to 6 GHz
<0.010 dB rms
Reference level
Range
±500 dB
Resolution
0.001 dB
Stability
2
300 kHz to 3 GHz
0.02 dB/1° C
3 GHz to 6 GHz
0.04 dB/1° C
Phase Characteristics
Dynamic accuracy (10 Hz IF BW)
Range
±180°
Display resolution
0.01°/division
Marker resolution
3
0.01°
Trace noise
(0 dBm at transmission port or +5 dBm at reflection
port, ratio measurement, 3 kHz BW)
300 kHz to 3 GHz
<0.038° rms
3 GHz to 6 GHz
<0.070° rms
Reference level
Range
-180° to +180°
Resolution
0.01°
Stability
30 kHz to 3 GHz
0.05°/° C
3 GHz to 6 GHz
0.20°/° C
Polar characteristics
Range
10 x 10
–12
to 1000 units full
scale
Reference
±500 units
1. See the Agilent 8753ET/ES descriptions and options for a functional description.
Measurement accuracy is dependent on the stability of the input signal.
2. Typical performance.
3. Marker resolution for magnitude, phase and delay is dependent upon measured