Agilent 8753ES (50-ohm systems)
Type-N test ports
Configuration
Network analyzer
8753ES
Standard and Option 006
Calibration kit
85032B
Test-port cables
11857D
System dynamic range
30 kHz to 50 kHz
70 dB
1
50 kHz to 300 kHz
90 dB
1
300 kHz to 16 MHz
100 dB
16 MHz to 3 GHz
110 dB
Option 014
108 dB
3 GHz to 6 GHz
105 dB
Option 014
103 dB
Measurement Port Characteristics
Frequency Range
Corrected
30 kHz–
300 kHz–
1.3 GHz–
3 GHz–
300 kHz
2
1.3 GHz
3 GHz
6 GHz
Directivity
50 dB
50 dB
47 dB
40 dB
Source match 49 dB
42 dB
36 dB
31 dB
Load match
50 dB
50 dB
47 dB
40 dB
Reflection
±(0.005 dB
±(0.009 dB
±(0.019 dB
±(0.070 dB
tracking
+0.02 dB/°C) +0.01 dB/°C) +0.02 dB/°C) +0.03 dB/°C)
Transmission ±(0.014 dB
±(0.013 dB
±(0.026 dB
±(0.065 dB
tracking
+0.02 dB/°C) +0.01 dB/°C) +0.02 dB/°C) +0.03 dB/°C)
Transmission uncertainty
1. Typical below 300 kHz.
2. Typical performance.
Magnitude
Phase
Magnitude
Phase
Reflection uncertainty