Accessories – Agilent Technologies 8753ET User Manual
Page 24

24
Calibration Kits
Vector accuracy enhancement procedures require
that the systematic errors of the measurement sys-
tem be characterized by measuring known devices
(standards) on the system over the frequency
range of interest. Return loss specifications or
typical values are provided where available for
the terminations and adapters.
85031B 7-mm calibration kit
Contains precision 7-mm standards used to cali-
brate the 8753ES for measurement of devices with
precision 7-mm connectors.
Includes
Part number
7 mm short/open circuit
85031-60001
7 mm 50-ohm terminations:
00909-60008
(two each)
Specifications for terminations
DC to 5 GHz
RL
≥
52 dB
5 to 6 GHz
RL
≥
46 dB
85032B 50-ohm type-N calibration kit
Contains precision 50-ohm type-N standards
used to calibrate the 8753ET/ES and 50-ohm test
sets for measurement of devices with 50-ohm
type-N connectors. Precision phase-matched 7-mm
to type-N adapters are included for accurate meas-
urements of non-insertable devices. Option 001
deletes the 7-mm to type-N adapters, which are
not needed with the 8753ET.
Includes
Part number
N-male 50-ohm termination
00909-60009
N-female 50-ohm termination
00909-60010
N-male short circuit
85032-60008
N-female short circuit
85032-60009
N-female open circuit
85032-60012
N-male open circuit
85032-60007
7-mm to N-male adapter
85054-60009
(two each)
7-mm to N-female adapter
85054-60001
(two each)
Specifications for terminations:
DC to 3 GHz
RL
≥
49 dB
2 to 3 GHz
RL
≥
46 dB
3 to 6 GHz
RL
≥
40 dB
Typical adapter characteristics:
DC to 6 GHz
RL
≥
30 dB
85033D 3.5-mm calibration kit
Contains a set of precision 3.5-mm standards
to calibrate the 8753ES and 50-ohm test sets for
the measurement of devices with precision 3.5-mm
and SMA connectors. Precision phase-matched
7-mm to 3.5-mm adapters are included for accurate
measurements of non-insertable devices. Option
001 deletes the 7-mm to type-N adapters.
Includes
Part number
3.5-mm-male 50-ohm termination 85033-60009
3.5-mm-female 50-ohm termination 85033-60010
3.5-mm-female short
85033-60014
3.5-mm-male short
85033-60013
3.5-mm-female open
85033-60012
3.5-mm-male open
85033-60011
7-mm to 3.5-mm female adapter
1250-1747
(two)
7-mm to 3.5-mm male adapter
1250-1746
(two)
Specifications for terminations:
DC to 1.3 GHz
RL
≥
46 dB
1.3 to 3 GHz
RL
≥
44 dB
3 to 6 GHz
RL
≥
38 dB
Typical adapter characteristics:
DC to 6 GHz
RL
≥
34 dB
85036B 75-ohm type-N calibration kit
Contains a set of precision 75-ohm type-N stan-
dards to calibrate the 8753ES with a 75-ohm test
set for measurement of devices with 75-ohm type-N
connectors. Precision phased-matched adapters
are included for accurate measurements of non-
insertable devices.
Includes
Part number
N-male 75-ohm termination
00909-60019
N-female 75-ohm termination
00909-60020
N-female 75-ohm short
85036-60011
N-male 75-ohm short
85036-60012
N-female open
85032-20001
N-male open
85032-60007
N-male to N-male 75-ohm adapter 85036-60013
N-female to N-female 75-ohm
85036-60014
adapter
N-male to N-female 75-ohm
85036-60015
adapter
Specifications for terminations:
DC to 2 GHz
RL
≥
46 dB
2 to 3 GHz
RL
≥
40 dB
Accessories