Agilent Technologies 8753ET User Manual
Page 9

9
1. One-port or enhanced-response calibration.
2. Enhanced-response or response-only calibration.
3. 2 dB less with Option 004.
4. Option 004 may degrade uncorrected source
match as much as 2 dB.
Transmission uncertainty
(enhanced-response calibration)
Transmission uncertainty
(response calibration)
Reflection uncertainty
(one-port calibration)
Agilent 8753ET (50-ohm systems)
Type-N test ports
Configuration
Network analyzer
8753ET Standard,
Option 006, or Option 004
Calibration kit
85032B
Test port cable
Part number 8120-4781
System dynamic range
300 kHz to 16 MHz
100 dB
16 MHz to 3 GHz
110 dB
3 GHz to 6 GHz
105 dB
Measurement Port Characteristics
Frequency Range
Uncorrected
300 kHz–1.3 GHz 1.3 GHz–3.0 GHz
3.0 GHz–6.0 GHz
Directivity
30 dB
24 dB
19 dB
Source match
4
25 dB
20 dB
14 dB
Load match
24 dB
19 dB
16 dB
Reflection
±1.0 dB
±1.0 dB
±2.0 dB
tracking
Transmission
±1.5 dB
±1.5 dB
±2.5 dB
tracking
Crosstalk
100 dB
100 dB
90 dB
Measurement Port Characteristics (continued)
Corrected
Reflection
measurements
1
300 kHz–1.3 GHz 1.3 GHz–3.0 GHz
3.0 GHz–6.0 GHz
Directivity
50 dB
47 dB
40 dB
Source match
42 dB
36 dB
31 dB
Load match
One-port cal
24 dB
19 dB
16 dB
Enhanced-
24 dB
19 dB
16 dB
reflection cal
Reflection
±(0.009 dB
±(0.019 dB
±(0.07 dB
tracking
+0.01 dB/°C)
+0.02 dB/°C)
+0.03 dB/°C)
Transmission
measurements
2
Source match
Enhanced-
42 dB
36 dB
31 dB
response cal
Response-only cal
3
25 dB
20 dB
14 dB
Transmission tracking
Enhanced-
±(0.006 dB
±(0.018 dB
±(0.054 dB
response cal
+0.01 dB/°C)
+0.02 dB/°C)
+0.03 dB/°C)
Response-only cal ±(0.033 dB
±(0.1 dB
±(0.27 dB
+0.01 dB/°C)
+0.02 dB/°C)
+0.03 dB/°C)