ACQUISITION - INTERLEAVED, ILVD
The INTERLEAVED command enables or disables Equivalent Time Mode for timebase settings where
both single shot and RIS mode are available.
Equivalent Time Mode is not available for sequence mode acquisitions. If sequence mode is on, ILVD
ON turns it off.
The response to the INTERLEAVED? query indicates whether the oscilloscope is in Equivalent Time
Mode.
Command Syntax
InterLeaVeD <mode>
<mode>:= {ON, OFF}
Query Syntax
InterLeaVeD?
Response Format
InterLeaVeD <mode>
Example
The following instructs the oscilloscope to use Equivalent Time Mode:
ILVD ON
Related Commands
TIME_DIV, TRIG_MODE