Response format, Example, Related commands – Teledyne LeCroy WaveAce 1000_2000 Remote Control User Manual
Page 43: Operator's manual

Operator's Manual
Parameter
Definition
MAX
Maximum value
MEAN
Mean value
MIN
Minimum value
NDUTY
- Duty cycle
NWID
Negative width (negative edge to positive edge)
OVSN
Overshoot negative/FOV
OVSP
Overshoot positive/ROV
PER
Period
PKPK
Peak to peak
PWID
Positive width (positive edge to negative edge)
RISE
Rise time 10% to 90%
RMS
Root mean square
RPRE
Rising edge preshoot
TOP
Top
WID
Width
Response Format
<trace>:PArameter_VAlue <parameter>,<value>,<state>...
<value>:=
A decimal numeric value
<state>:= {OK, AV, PT, IV, NP, GT, LT, OF, UF, OU}
NOTE: If <parameter> is not specified, or is equal to ALL, all standard voltage and time parameters are returned
followed by their values and states. When PAVA? Is used to query a Custom parameter, the <trace> prefix is
returned for consistency. The source for the measurement is configured using the PACU command.
Example
The following instruction query reads the rise time of Trace B (TB):
TB:PAVA? RISE
Response message:
TB:PAVA RISE,3.6E-9S,OK
Related Commands
CURSOR_MEASURE, CURSOR_SET, PARAMETER_CUSTOM
920836 Rev B
41