Acquisition - trig_select, trse, Command syntax, Query syntax – Teledyne LeCroy WaveAce 1000_2000 Remote Control User Manual
Page 33: Response format, Example

Operator's Manual
ACQUISITION - TRIG_SELECT, TRSE
The TRIG_SELECT command selects the condition that triggers the acquisition of waveforms. Only the
Edge trigger is supported.
Command Syntax
TRig_Select EDGE,SR,<source>,HT,TI,HV,<hold_value>
<source>:= {C1,C2,C3,C4,LINE,EX,EX5}
<hold_value>:=
Refer to the trigger information provided for your oscilloscope model.
NOTE: When specifying <hold_value>, the unit S (seconds) is optional.
Query Syntax
TRig_SElect?
Response Format
TRig_SElect <trig_type>,SR,<source>,HT,<hold_type>,HV,<hold_value>
HV2 only returned if <hold_type> is P2 or I2.
Example
The following instruction selects the single-source trigger with Channel 1 as trigger source. Hold type
and hold value are chosen as "pulse smaller" than 20 ns:
TRSE SNG,SR,C1,HT,PS,HV,20 NS
920836 Rev B
31