Teledyne LeCroy WaveRunner 6 Zi and 12-Bit HRO Getting Started Manual User Manual
Page 126

WaveRunner 6 Zi and 12-Bit HRO
118
922136-00 Rev A
the Summary box of the Qx dialog as follows (shown on the Dual Param
Compare right-hand dialog):
1. Touch Analysis → Pass/Fail Setup... from the menu bar.
2. Touch one of the Qx tabs and a setup dialog for the position is
shown.
3. Touch inside the Condition field in the main dialog and select
DualParamCompare.
4. Touch inside the Source1 and Source2 fields and select a source
from the pop-up menu.
5. Touch inside the ParamCompare mini-dialog field and select a
source from the pop-up menu.
6. Touch inside the Compare Values field and select All or Any from
the pop-up menu.
By selecting All, the test are true only if every waveform falls within
the set limit. Selecting Any makes the test true if just one waveform
falls within the limit.
7. Touch inside the Condition field on the ParamCompare part of the
dialog and select a math operator from the pop-up menu.