Use the AED when a suspected cardiac arrest victim has an apparent

LACK OF CIRCULATION indicated by:

Unconsciousness and

Absence of normal breathing and

Absence of a pulse or signs of circulation.

When a victim is less than 8 years of age, or weighs less than 55 lbs
(25 kg), the ZOLL AED Plus should be used with ZOLL AED Plus
Pediatric Electrodes. Therapy should not be delayed to determine the
patient’s exact age or weight.

Do NOT use when a patient:

Is conscious; or

Is breathing; or

Has a detectable pulse or other signs of circulation.

The ZOLL AED Plus external defibrillator is intended to be used by
personnel who are qualified by training in the use of the AED Plus device
and basic life support, advanced life support, or other physician-
authorized emergency medical response to defibrillate victims of cardiac
arrest. The CPR monitoring function provides a metronome designed to
encourage rescuers to perform chest compressions at the AHA
recommended rate of 100 compressions per minute. Voice and visual
prompts encourage a minimum compression depth of 2 inches for adult
patients. The CPR monitoring function is not intended for use on patients
under 8 years of age.

After one shock...

Person collapses—
suspected
cardiac arrest.

Unit OK.

Turn unit on.

Self-test; unit OK.

Stay calm.

Call for help.

Attach defib pads to patient’s
bare chest.

No shock advised.

Don't touch patient.
Analyzing.

start

here

Indicator light flashes.

Indicator light flashes.

Call or send someone
for help.

Indicator light flashes.

Check responsiveness.

Shout, "Are you OK?”
as you shake the person.

Indicator light flashes.

Prompts you may see in the LCD screen:

Text messages
Shock counter
Elapsed time
ECG waveform

What you see.

What you hear.

What you do.

Start CPR.

Administer CPR:
2 breaths for every
30 compressions.

Indicator light flashes.

Other prompts you

may hear:

Plug in cable.

Unit failed.

Release shock button.

Change batteries.

(n) shocks delivered.

What you should do:

Attach electrode cable.

Turn unit off and then on.

If unit still fails, perform CPR.

Let go of shock button.

Replace all 10 batteries.

After 2 minutes of CPR...

Stop CPR.

LCD screen

Attach adult or infant/child
electrode pads to patient's
bare chest.

Place electrode pads as
shown on patients under
8 years of age or < 55 lbs (25 kg).

Follow electrode pads
package instructions.

Unit equipped to
treat adult and
pediatric patients.

©

30

Shock advised.
Don't touch patient.
Press flashing shock button.

Indicator light flashes.

Give

STAND CLEAR

Warning.

DO NOT touch patient.

Press button.

Shock delivered.

2

INDICATIONS FOR USE

CONTRAINDICATIONS FOR USE

INTENDED USER

ZOLL Medical Corporation

269 Mill Road

Chelmsford, MA 01824-4105

978-421-9655 • 800-348-9011

www.zoll.com