E Series - Pulse CO-Oximetry (SpO
2
, SpCO, and SpMet) Option Insert
14
9650-1211-01
Rev.
H
Note: The E Series Pulse Oximetry Option is calibrated for functional saturation.
Alarm Limits
On/Off displayed on monitor. User selectable.
SpO
2
: High 72 - 100% saturation, Low 70 - 98% saturation
SpCO: High 2 - 100% saturation, Low 0 - 99% saturation
SpMet: High 1 - 100% saturation, Low 0 - 99% saturation
Pulse Rate: High 60 - 235 beats per minute, Low 20 - 100 beats per minute
SpO
2
Wavelength for LNCS
Sensors
Nominal Red LED Wavelength: 660 nanometers
Nominal Infrared LED Wavelength: 905 nanometers
Energies (Radiant Power)
of light for LNCS Sensors
at 50 mA pulsed
15 mW
SpO
2
Wavelength for
Rainbow Sensors
The Rainbow sensors use 8 different LEDs with wavelengths
of 610 - 905 nanometers
Energies (Radiant Power)
of light for Rainbow Sensors
at 100 mA pulsed
25 mW
Bio-Compatibility
Patient contacting material meets requirements of ISO 10993-1, Biological
Evaluation of Medical Device - Part I, for external devices, intact surfaces and
short-term exposure
Environmental
Operating Temperature:
0º to 40º C (32º to 104º F)
Storage Temperature:
-40º to 70º C (-40º to 158º F)
Note: The E Series device may not perform to specifications when stored at the
upper or lower extreme limits of storage temperature and immediately put
into use
Electromagnetic Immunity
(SpO
2
Option Only)
AAMI DF-80; EN61000-4-3:2002 to 10 V/m
Operating Time
For a new, fully charged PD4410
battery pack at 20°C (68°F):
•
35 defibrillator discharges at
maximum energy (200 J), or
•
2.0 hours minimum of continuous
ECG monitoring, or
•
1.75 hours of continuous ECG
monitoring/pacing at 60 mA, 70
beats per minute.
For a new, fully charged lithium ion
battery pack at 20°C (68°F):
•
95 defibrillator discharges at
maximum energy (200 J), or
•
3.75 hours minimum of continuous
ECG monitoring, or
•
3.25 hours of continuous ECG
monitoring/pacing at 60 mA, 70
beats per minute.