Fully Automatic AED Plus Administrator’s Guide
19
Batteries produced by all recommended manufacturers have a 10 year from date of manufacture shelf
life when not installed in an Fully Automatic AED Plus unit.
The following examples demonstrate how to read date codes on Duracell, Sanyo and Varta batteries.
Duracell:
The date code is YYYY/MM.
Sanyo:
The first letter indicates the year of manufacture (Example: A=1996, B=1997, etc.).
The second letter indicates the month of manufacture (Example: A=January, B=February, etc.).
Varta:
The date code is MMYY.
To install the batteries:
1.
Ensure that the Fully Automatic AED Plus is turned off. Open the battery compartment by
removing the battery cover from the back of the unit.
To remove the battery cover, insert a small tool (e.g., flat blade screw driver) into the two slots on
the rear of the unit to depress the latches, then insert the tool into the groove on the bottom to lift
the cover (see Figure 7).
Figure 7: Removing the Battery Compartment Door
1) Depress latch
here
2) Lift cover here
Battery
Compartment
Door