Additional Messages and TroubleShooting

The following chart lists the messages that may appear on the M Series unit relating to the Biphasic Option, why the message
appeared, and the action(s) to take.

The operator should become thoroughly familiar with this information before using the device.

Message

Possible

Cause(s)

Recommended

Action(s)

200 J MAX BIPHASIC

Appears when trying to select an energy higher

No higher energy is available.

than

200

Joules.

Use

200J

setting.


BRIDGE TEST FAILED

Biphasic module not operating properly while

Try charging again. Attempt to

charging.

clear the message by turning the

Selector

Switch

to

OFF

then

back

to

the

desired

operating

mode.

If

Fault

persists

contact

ZOLL

Technical

Service

Department.


BRIDGE SHORT

Current higher than expected was detected.

Ensure pads/paddles are used

properly.

Attempt

to

clear

the

message

by

turning

the

Selector

Switch to OFF then back to the

desired

operating

mode.

If

Fault

persists

contact

ZOLL

Technical

Service

Department.

Additional Specifications and Changes

General

Refer to M Series Operator’s Guide for ALL Specifications except the following:

Waveform:

Rectilinear Biphasic

Energy Settings:

1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 50, 75, 100, 120, 150, 200

Charge Time:

Less than 6 seconds with a new fully charged battery

(first 15 charges at 200J)

Messages:

“200 J MAX BIPHASIC”, “BRIDGE TEST FAILED”, “BRIDGE SHORT”

Operating

Time:

For a new, fully charged battery pack at 20°C: 40 defibrillator

discharges

at

maximum

energy (200J), or 2.75 hours minimum of

continuous

ECG

monitoring, or 2.25 hours of continuous ECG

monitoring/pacing at 60 mA, 80 beats/min.