Introduction, Conventions, Safety – ZOLL Propaq MD Rev A User Manual

Page 3: Warnings

Advertising
background image

9650-0809-01 Rev. A

Propaq ® MD Battery Pack

Guide

1

Introduction

The Propaq

®

MD Battery Pack is a rechargable, sealed lithium ion battery for use in

compatible ZOLL

®

defibrillators and chargers.

It is important that you read and understand the information in this manual. The proper use of
the Propaq MD Battery Pack is critical to the reliable operation of your ZOLL defibrillator.

Conventions

This manual uses the following conventions:

Warning statements describe conditions or actions that can result in personal injury or
death.

Caution

Caution statements describe conditions or actions that can result in damage to the Propaq MD

Battery Pack.

Note: Notes contain additional information on using and maintaining the Propaq MD Battery

Pack.

The manual displays button and LED labels in bold face type.

Safety

The following sections provide important warnings, cautions, and notes regarding the ZOLL
Propaq MD Battery Pack.

Warnings

Locate ZOLL Propaq MD units where they cannot harm the patient or medical personnel if

they fall -- accidental sliding or drop of the Propaq MD Battery Pack may cause physical

injury. Damaged batteries may expose users to a safety hazard. Should the Propaq MD

Battery Pack be dropped, inspect and test the unit prior to further use following the

procedures described in the Maintenance section of this Guide.

Avoid prolonged exposure of the Propaq MD Battery Pack to direct sunlight to minimize

heating and damage that may occur from exposure to ultraviolet light.

Propaq MD Battery Packs have been tested against interference from radio frequency

emissions typical of two-way radios used in emergency service/public safety activities.

Refer to the Propaq MD Operator’s Guide to determine recommended operating distances

from RF transmission equipment

.

Do not use Propaq Battery Packs unless they have been visually inspected, tested, and

recalibrated by appropriate personnel and you have been adequately trained in the proper

use of these batteries. See the Maintenance section of this guide for more information.

Do not disassemble the Propaq MD Battery Pack.

Charge the Propaq MD Battery Pack in a well-ventilated area.

You must perform periodic inspection, capacity testing, and functional testing of

rechargable defibrillator batteries to ensure the availability of safe and reliable batteries. See

the Maintenance section of this guide for more information.

If the battery in use on a Propaq MD unit indicates a fault condition or any operating state

that can prevent the effective use of the defibrillator, replace that battery as soon as possible.