Cleaning, Maintenance – ZOLL SurePower Rev B Battery Pack User Manual

Page 7

Advertising
background image

9650-0536-01 Rev. B

ZOLL SurePower Battery Pack Guide

5

Cleaning

Clean the SurePower Battery Pack with a soft damp cloth and mild soap.

Do not immerse or set the SurePower Battery Pack in fluids of any kind. Do not use alcohol or
ketones to clean the SurePower Battery Pack.

Maintenance

Perform a battery check, which verifies that the battery is fully charged and fault free, as part of
any regularly scheduled defibrillator test. Press the battery’s Display button to perform the
battery check.

As part of your defibrillator maintenance program, verify that the battery’s LEDs work (press
the Display button to test LEDs) -- if the LEDs do not work, remove the battery from service.
Also check the battery casing for cracks or other damage -- if the battery is cracked or damaged,
remove it from service.

SurePower Battery Pack Icon

LED Indications

This is the Recalibration icon.

This LED lights for approximately 10 seconds (after you
press and release the Display button) if the battery needs to
be calibrated. If the Recalibration LED lights, the runtime
indicator will not display run time for that battery. For best
performance of the battery, you should recalibrate the
battery as soon as possible.

To manually recalibrate the SurePower Battery Pack, you
can insert the battery into the SurePower

Charger Station

and perform a ManualTest (for more information, see the
ZOLL SurePower Charger Station Operator’s Guide

).

After you recalibrate the battery, the Recalibration LED will
only flash when you press the Display button.

This is the Fault Icon.

If the SurePower Battery Pack detects a battery fault, this
LED lights for approximately 10 seconds after you press the
Display

button.

If the Fault LED lights, the runtime indicator will not display
run time for that battery.

See the following section for more information on battery
faults.