Yokogawa EJX115A User Manual

Page 36

Advertising
background image

<5. Parameter Summary>

5-4

IM 01C25T03-01E

No. Parameter name

Item

*

1

R/W

Content

Default value

Applicable

model

U

pl

oa

d

da

ta

F

P

L

J

ADJUST

Adjusting data

J09 ADJ UNIT

Pressure adjusting unit select

W % or PRES UNIT

PRES UNIT

J10 ADJ PRES

Adjustment reference pressure

R

Unit specified in J09

J11 P ZERO ADJ

Automatic zero adjustment

W -32000 to 32000, unit

specified in J09

0.00000 kPa

J12 P SPAN ADJ

Automatic span adjustment

W -32000 to 32000, unit

specified in J09

100.000 kPa

J15 P ZERO DEV

Manual zero adjustment

W -32000 to 32000, unit

specified in J09

100.000 kPa

J16 P SPAN DEV

Manual span adjustment

W -32000 to 32000, unit

specified in J09

100.000 kPa

J20 ADJ SP

Adjustment reference pressure

of SP

R

Unit specified in J09

— 

J21 SP ZERO ADJ

Automatic SP zero adjustment

W -32000 to 32000, unit

specified in J09

0.00000 MPa

— 

J22 SP SPAN ADJ

Automatic SP span

adjustment*

2

W -32000 to 32000, unit

specified in J09

— 

J25 SP ZERO DEV

Manual SP zero adjustment

W -32000 to 32000, unit

specified in J09

0.00000 MPa

— 

J26 SP SPAN DEV

Manual SP span adjustment

W -32000 to 32000, unit

specified in J09

0.00000 MPa

— 

J40 OUTPUT 4mA

4 mA adjustment

W -10.000 to 10.000%

0.000%

J41 OUTPUT 20mA

20 mA adjustment

W -10.000 to 10.000%

0.000%

J45 AMP TEMP

Amplifier temperature

R

Unit specified D40

0 degC

J50 ADJ WHO

Adjustment information

W 8 alphanumeric characters

J51 ADJ DATE

Adjustment information

W 16 alphanumeric characters

J52 ADJ LOC

Adjustment information

W 8 alphanumeric characters

J53 ADJ DESC

Adjustment information

W 16 alphanumeric characters

J55 EXT ZERO ADJ

External zeroing permission

W INHIBIT or ENABLE

ENABLE

J56 CLEAR ADJ

Clear adjustment

W ---, PRES, SP, 4-20mA, or

ALL

---

J60 SELF CHECK

Self-diagnostics

R

See A60

K

TEST

Test parameters

K10 OUTPUT X %

Test output % setting

W Within a range between D20

and D21

0.00%

K40 DO TEST

Test contact output

W OFF or ON

OFF

K45 TEST TIME

“OUTPUT X %” and “DO

TEST” duration time selection

W 10 min, 30 min, 60 min,

3 hour, 6 hour, 12 hour

10 min

K50 TEST KEY1

Special maintenance

parameter

W

K51 TEST KEY2

Special maintenance

parameter

W

K52 TEST KEY3

Special maintenance

parameter

W

K53 TEST KEY4

Special maintenance

parameter

W

K60 SELF CHECK

Self-diagnostics

R

See A60

M

DEVICE INFO

Device information

M10 SERIAL NO.

Serial number

R

M11 MFTR. DATE

Manufactured date

R

M12 EXTRA NO.

Customization number

R

M15 SOFT REV

Software revision

R

M16 BRAIN REV

BRAIN protocol revision

R

M17 MEMO1

Memo

W 16 alphanumeric characters

M18 MEMO2

Memo

W 16 alphanumeric characters

M19 MEMO3

Memo

W 16 alphanumeric characters

M20 ISOL MATL

Capsule material

W

M21 FILL FLUID

Fill fluid

W

*1: R/W: R = Read only, W = Read & Write

*2:

The default value shows MWP (Maximum working pressure) of the capsule.

Since the working pressure limit varies according to the Model, refer to the General Specifications section in each user’s manual.

Advertising