4 printout (for bt200 printer option), 1) printout of all parameters, 2) printout by menu item – Yokogawa EJX115A User Manual

Page 11

Advertising
background image

<3. Operation>

3-3

IM 01C25T03-01E

3.1.3 Calling Up Menu Addresses Using the Operating Keys

DATA

DIAG

PRNT

ESC

PARAM

A60:SELF CHECK

GOOD

CODE

CAPS

CLR

ESC

--WELCOME--

BRAIN TERMINAL

ID: BT200
check connection

push ENTER key

UTIL

FEED

esc

PARAM

01:MODEL

EJX110A-DM

02:TAG NO.

YOKOGAWA

03:SELF CHECK

GOOD

OK

STARTUP
SCREEN

UTILITY SCREEN

FUNCTION SCREEN

MENU
SCREEN

SETUP
SCREEN

INITIAL
DATA
SCREEN

PARA-
METER
SCREEN

(HOME MENU SCREEN)

(SET MENU SCREEN)

(ADJ MENU SCREEN)

See the BT200 user’s manual for details concerning uploading

and downloading parameters and printouts (BT200-P00).

(UTIL)

(SET)

(ESC)

HOME

SET

ADJ

(ADJ)

The utility screen contains the

following items.

1. BT200 ID settings

2. Security code settings

3. Switching language of messages

(Japanese or English)

4. LCD contrast setting

5. Adjusting printout tone

(BT200-P00 only)

FUNC

1.MENU

2.UPLOAD TO HHT

3.DOWNLOAD TO INST

4.PRINT ALL DATA

HOME

SET

ADJ

ESC

MENU

A:DISPLAY

B:SENSOR TYPE

C:BASIC SETUP

D:AUX SET1

E:AUX SET2

G:ALARM SET

HOME

SET

ADJ

ESC

MENU

C:BASIC SETUP

D:AUX SET 1

E:AUX SET 2

G:ALARM SET

H:AUTO SET

I:DISP SET

HOME

SET

ADJ

ESC

MENU

J:ADJUST

K:TEST

M:DEVICE INFO

P:RECORD

T:CHARACTERIZR

DATA

DIAG

PRNT

ESC

DATA

DIAG

PRNT

ESC

DATA

DIAG

PRNT

ESC

PARAM

C60:SELF CHECK

GOOD

PARAM

C22:PRES URV

100.00 kPa

C23:PRES POINT

DATA

DIAG

PRNT

ESC

DATA

DIAG

PRNT

ESC

PARAM

J60:SELF CHECK

GOOD

PARAM

J12:P SPAN ADJ

100.000 %

J15:P ZERO DEV

DATA

DIAG

PRNT

ESC

PARAM

J09:ADJ UNIT

%

J10:ADJ PRES

0.02563 %

J11:P ZERO ADJ

0.04792 %

SET

C10:TAG NO.

YOKOGAWA

YOKOGAWA

F0309.ai

DATA

DIAG

PRNT

ESC

PARAM

A10:OUTPUT

0.0 %

A11:PRES

0.0000 kPa

A15:OUTPUT mA

4.000 mA

DATA

DIAG

PRNT

ESC

PARAM

C10:TAG NO.

YOKOGAWA

C20:PRES UNIT

kPa

C21:PRES LRV

0.00000 kPa

PARAM

A16:ENGR.OUTPUT

0.0000 mmHg

A17:ENGR.EXP

UTILITY

1.ID

2.SECURITY CODE

3.LANGUAGE SELECT

4.LCD CONTRAST

5.PRINTER ADJUST

3.1.4 Printout (for BT200 printer option)

(1) Printout of All Parameters

Select 4. PRINT ALL DATA from the function screen

to output a list of all parameters. It takes about

10 minutes to complete the printout.

(2) Printout by Menu Item

To printout the parameters for a specific screen,

push the function key corresponding to screen’s

PRNT.

Advertising