Important – Yokogawa EJA310E User Manual

Page 39

Advertising
background image

<7. Operation>

7-6

IM 01C25B01-01E

IMPORTANT

• Do not turn off the power to the transmitter

immediately after completion of the change

in the LRV and/or URV setting(s). Note

that powering off within thirty seconds after

setting will cause a return to the previous

settings.

• Changing LRV automatically changes URV

to the following value.

URV=previous URV+(new LRV–previous LRV)
• If the range-setting push-button and external

zero-adjustment screw are not touched

during a range-change operation, the

transmitter automatically switches back to

the normal operation mode.

F0708.ai

Integral indicator

Range-setting switch

(Push-button)

Note 1: Use a thin bar which has a

blunt tip, e.g., a hexagonal

wrench, to press the

range-setting push-button.

Note 2: The push-button is located

in either lower right or

lower left portion of the

LCD indicator.

Figure 7.7

Range-setting Switch

Advertising
This manual is related to the following products: