Yokogawa In-Situ Gas Analyzer TDLS200 User Manual

Page 23

Advertising
background image

23

IM11Y01B01-12E-A

1st Edition


Once visual alignment is achieved on both the launch and detect and clean, dry purge gas
has been flowing, power may be applied to the analyzer.

Begin the alignment process by starting at the launch end; maximize the signal (using the
transmission signal and value) by moving the launch around using the alignment bellows.

Once maximum signal is achieved, move to the detect end and using the method described
in the previous slide, move the LAO/detect assembly around, until maximum signal is
achieved. (Note: CO detectors are more sensitive to miss-alignment than the O2 detector)