Adjustment of Detector Gain after Optimization of Alignment –
Please read these directions carefully before attempting work and contact your local Yokogawa
service center if there is any doubt about how to perform this procedure.
Alternatively, Yokogawa service personnel can perform this procedure – please contact you local
service center for details, rates, availability, etc.

Caution – Electrostatic Sensitive!

INTRODUCTION

For TDLS200 measurement over long optical path length, the laser beam is usually configured with
a small diverging angle (either with or without large aperture detector lens). In factory, the detector
signal gain is not optimized for the real installation. During the analyzer field start up service, the
detector gain needs to be adjusted to achieve the best analyzer performance.

This document gives the guidance and procedures to adjust the detector board gain for services
and customers after the analyzer is installed and powered up.

This operation should be done by qualified personnel. Read instructions fully before starting this
operation. Any doubts or questions, please contact Yokogawa.

PHOTOS OF DETECTOR BOARD

RESISTOR KIT BAG

For a TDLS200 analyzer with diverging laser beam configuration (either with or without large
aperture detector lens), a resistor bag is attached inside the detector box. The resistor bag contains
the resistors with the following values. All of them are ¼ W, 5% tolerance metal film through-hole
resistors.

Resistor Bag for Analyzers With Diverging Beam Optics

Resistance (Ω)

100

1k

1.5k

2.2k

3.3k

5.1k

7.5k

11k

15.8k

24.3k

Quantity

1

2

2

2

2

2

2

2

2

2

Low-temp detector board (< 600°C)

High-temp detector board (> 600°C)