3 installation and wiring> 3-20 – Yokogawa In-Situ Gas Analyzer TDLS200 User Manual

Page 33

Advertising
background image

3.11.1 Purging Analyzer for Hazardous Areas (with On-Line Validation)

• NEC/CSA Class 1, Division 2, Groups A-D

ATEX Zone 2 CAT 3 Dual regulators must be used on the inlet!

The block diagram below shows the sections of the analyzer that require nitrogen purging. A Z-Type purge control
system is fi tted the Launch Unit and it includes a local indicator (Bright Green, rugged light) and pressure switch
alarm contacts (open on loss of purge pressure). The purging should be carried in sequence typically as shown
below. All purge gas connections are ¼” od Tube fi ttings.

3.11.2 Purging Analyzer for Hazardous Areas (without On-Line Validation)

• NEC/CSA Class 1, Division 2, Groups A-D
• ATEX Zone 2 CAT 3 – Dual regulators must be used on the inlet!

The block diagram below shows the sections of the analyzer that require nitrogen purging. A Z-Type purge
control system is fi tted the Launch Unit and it includes a local indicator (Bright Green, rugged light) and pres-
sure switch alarm contacts (open on loss of purge pressure).

Figure 26 - Purge Flow Diagram when not using on line validation

Main

Electronic

Housing &

Purge

System

Laser

Module

Check Gas

Flow Cell

P

roce

ss

Int

erf

a

c

e

P

roce

ss

Int
erf
a

c

e

Detect

Module

Detect

Electronic

Housing

Nitrogen or I/A Purge Gas

Main

Electric

Housing

& Purge

System

Detect

Electronic

Housing

Main

Electronic

Housing &

Purge

System

Laser

Module

Check Gas

Flow Cell

Proce

s

s

Inte

rfa

ce

Proce

s

s

Inte

rfa
ce

Detect

Module

Detect

Electronic

Housing

Nitrogen or I/A
Purge Gas

Nitrogen or I/A
Purge Gas

Figure 25 - Purge Flow Diagram when using on line validation

Main

Electric

Housing

& Purge

System

Detect

Electronic

Housing

Nitrogen or I/A Purge
Gas

Check Gas

Flow Cell

<3 INSTALLATION AND WIRING> 3-20

IM 11Y01B02-11E-A 5th Edition :June 5, 2012-00