3 installation and wiring – Yokogawa In-Situ Gas Analyzer TDLS200 User Manual

Page 34

Advertising
background image

3-21

3.11.3 Purging Analyzer and Universal Power Supply and/or URD for Hazardous
Areas (with On-Line Validation)

• NEC/CSA Class 1, Division 2, Groups A-D

• ATEX Zone 2 CAT 3 – Dual regulators must be used on the inlet!

The block diagram below shows the sections of the analyzer that require nitrogen purging when using in
conjunction with either or the Universal Power Supply and Universal Remote Display. AZ-Type purge control
system is fi tted the Launch Unit and it includes a local indicator (Bright Green, rugged light) and pressure switch
alarm contacts (open on loss of purge pressure).

3.11.4 Purging Analyzer and Universal Power Supply and/or URD (not using On-Line Validation)


• NEC/CSA Class 1, Division 2, Groups A-D

• ATEX Zone 2 CAT 3 – Dual regulators must be used on the inlet!

The block diagram below shows the sections of the analyzer that require nitrogen purging. A Z-Type purge
control system is fi tted the Launch Unit and it includes a local indicator (Bright Green, rugged light) and
pressure switch alarm contacts (open on loss of purge pressure). The purging should be carried in sequence
typically as shown below. All purge gas connections are ¼” od Tube fi ttings.

Main

Electronic

Housing &

Purge

System

Laser

Module

Check Gas

Flow Cell

Proce

s

s

Inte

rfa

ce

Proce

s

s

Inte

rfa
ce

Detect

Module

Detect

Electronic

Housing

Nitrogen or I/A Purge
Gas

Nitrogen or I/A Purge
Gas

Power

Supply

or URD

Figure 27 - Purge Flow Diagram when using on line validation

Nitrogen or I/A Purge
Gas

Check Gas
Flow Cell

Detect
Electronic
Housing

Main

Electronic

Housing &

Purge

System

Laser

Module

Check Gas

Flow Cell

Process

Interface

Process

Interface

Detect

Module

Detect

Electronic

Housing

Nitrogen or I/A Purge Gas

Power

Supply

or URD

Figure 28 - Purging Analyzer for Hazardous Areas -NO- On-Line Validation

Main
Electronic
Housing
& Pureg
System

Detect
Electronic
Housing

Power
Supply
or URD

Check Gas
Flow Cell

<3 INSTALLATION AND WIRING>