Chapter 2 troubleshooting, 1 failure analysis flow chart, 1 failure analysis flow chart -1 – Yokogawa Button Operated MV2000 User Manual

Page 11

Advertising
background image

2-1

Troubleshooting

SM MV1000-01E

1

2

3

4

5

6

7

2.1 Failure Analysis Flow Chart

When a failure occurs, refer to the flow chart below for corrective actions.

See No. 5 of the checklist

in section 2.2.

See No. 1 of the checklist in

section 2.2.

Has the backlight gone out due

to the LCD’s backlight saver?

Other than the display,

is the instrument functioning

normally?

Is an appropriate power

supply being input?

Is the power supply wired

correctly?

Is the power switch* ON?

Does the switch operate

normally?

Doesn’t function at all

(nothing is displayed)

Turn the switch ON correctly.

Wire it correctly.

Supply power of voltage and

frequency that conforms to

the specifications.

Restore the original brightness

through key operation.

Try replacing the backlight.

YES

NO

YES

NO

YES

NO

YES

NO

NO

YES

NO

YES

Has the instrument

been operated longer than the

recommended replacement period

for the backlight (5 years)?

Chapter 2

Troubleshooting

Advertising
This manual is related to the following products: