SM MV1000-01E

7.2 Recommended Replacement Periods for Worn

Parts

To preserve the reliability of the MV and to use the MV in a good condition for an
extended time, it is recommended that periodic replacements be made on parts. The
replacement parts may change to accommodate preventive maintenance over extended
time. Be sure to check with your nearest YOKOGAWA dealer.
The following table shows the recommended replacement period for expendable parts.
The replacement period shown here applies when the MV is used under standard
operating conditions. For the actual replacement period, consider the actual conditions
of use. Replacement of parts will be carried out by a YOKOGAWA engineer or an
engineer certified by YOKOGAWA. Contact your nearest YOKOGAWA dealer when such
replacement is necessary.

MV1000

Item

Replacement

period

Name

Part No.

Quantity

Used

Notes

LCD

5 years

Bezel Assembly

B8805BA

1

Battery

10 years

Battery Assembly

B9900BR

1

Aluminum

electrolytic

capacitor

5 years*

Power supply Assembly –

1

5 years*

AD Assembly

Depends on the model

* Replacement Period at the Upper Limit of the Normal Operating Temperature (40°C)
The replacement period varies depending on the temperature in which the instrument

is operated, and the instrument’s specifications. If the instrument is used in a 30°C
environment, it may be operational for 10 years or more.

MV2000

Item

Replacement

period

Name

Part No.

Quantity

Used

Notes

LCD

5 years

Bezel Assembly

B8806BA

1

For English

B8806BB

1

For Japanese

Battery

10 years

Battery Assembly

B9900BR

1

Aluminum

electrolytic

capacitor

5 years*

Power supply Assembly –

1

5 years*

AD Assembly

Depends on the model

* Replacement Period at the Upper Limit of the Normal Operating Temperature (40°C)
The replacement period varies depending on the temperature in which the instrument

is operated, and the instrument’s specifications. If the instrument is used in a 30°C
environment, it may be operational for 10 years or more.

Note

• The LCD should be replaced when the brightness of the LCD is reduced by half its original

brightness. The brighter the display, the shorter the “half-life” of the LCD. Reductions in
brightness vary depending on operating conditions, and is a subjective determination. The
actual replacement period should be determined based on these considerations.

• The LCD screen may start to yellow over time. The brighter the screen, the faster such

changes may progress.