Withstand voltage, Withstand voltage -5 – Yokogawa Button Operated MV2000 User Manual

Page 20

Advertising
background image

3-5

SM MV1000-01E

Testing

1

2

3

4

5

6

7

MV2000

Power inlet

Functional ground

terminal

Power supply

screw terminals

Protective ground

terminal

Withstand Voltage

● Test Procedures

Perform the test with a withstanding voltage tester, and check whether the results

satisfy the reference values. Perform the measurement with the power switch turned
ON.

Test time: 1 minute

Tested Terminals

Applied

Voltage

Maximum Allowable
Leakage Current

Condition

Between the power and

protective ground terminals

2.3 kVAC

10 mA

For models with 100-240 VAC

power supply.

Short the L and N terminals.

0.5 KVAC

10 mA

For models with 12 VDC power

supply.

Short the + and – terminals.

Between the input and

protective ground terminals

1.5 kVAC

2 mA

Short all input terminals.

Between input terminals

1.0 kVAC

1 mA

Short the odd and even channels

of the A/+, B/- terminals*, and

measure between odd and even

channels.

Between relay contact

output and protective

ground terminals

1.6 kVAC

2 mA

With the /A1, /A2, /A3, /A4, or

/F1 option. Short all relay contact

output terminals.

Between the remote control

input and protective ground

terminals

1.0 kVDC

2 mA

With the /R1 option. Short all

remote control input terminals.

Between pulse input and

protective ground terminals

1.0 kVDC

2 mA

With the /PM1 option. Short all

pulse input terminals.

Between 24 V transmitter

power supply output and

protective ground terminals

500 VAC

10 mA

With the /TPS2 or /TPS4 option.

Short all 24 V transmitter power

supply output terminals.

* The b terminals on the following models are independent on all channels. Short the odd

and even channels of the A/+, B/-, and b terminals, and measure between odd and even
channels.
• MV1004, MV1008, MV2008
• With the /N1 or /N2 option.

3.4 Test Procedures

Advertising
This manual is related to the following products: