3-3

SM 04P01B01-01E

Testing

3

3.3

Performance Tests

Read the warning and cautions below before beginning tests.

WARNING

• Circuit patterns of the printed circuit board are exposed. Be careful when

handling so that hands or fingers are not injured by the protruding pins.

CAUTION

• Circuit patterns of the printed circuit board are exposed. If other metallic

materials touch these patterns, electrical shorting will occur, causing the circuit
to be damaged or burnt.

• It is sometimes necessary to turn the recorder upside down for adjustment. Do

not drop the instrument, or allow it to fall over.

Test Environment

Item

Conditions

Ambient temperature

23 ± 5°C

Relative humidity

55 ± 20%

Atmospheric pressure

86 kPa-106 kPa

Power supply voltage

Nominal voltage ±1%

Power supply frequency

Nominal frequency ±1%

Power supply waveform

Distortion factor of 5% or less

DC power supply ripple

Content ratio of 0.1% or less

Vibration

A value having negligible effects on the instrument

Position

Within ±3% of the specified position

Interference*

A value that does not affect the measured results

Electrical field

A value that does not affect the measured results

Magnetic field

A value that does not affect the measured results

Atmospheric pollutants

Levels of corrosive gasses, vapors, salts, and dust that
do not affect the measured results.

Other external effects**

A value that does not affect the measured results

*

Interference

refers to common mode noise, series mode noise, power supply noise, and other

phenomena in the signal line.

** Other external effects include luminance, draft, ultrasonic waves, and radiation.

Test Instruments

Instrument

Specifications

DC voltage generator

Accuracy: ±0.01% of setting or better

DMM

Accuracy: ±0.01% of rdg or better

Variable resistors

Accuracy: 0.01% + 2 mΩ or better

Insulation tester

500 VDC

Withstand voltage tester

1 to 1.5 kVAC

Glass scale

105 mm or more

Advertising