Recording accuracy, Z-fold chart paper feed accuracy – Yokogawa µR20000 User Manual

Page 27

Advertising
background image

3-14

SM 04P01B01-01E

Recording Accuracy

Overview

Set the channels to the 2 V range and confirm that the recording accuracy is within
the criteria.
Pen Model
Check the accuracy of all pens. Use standard pens (or equivalent).

*

A “standard pen” is one whose tip does not fluctuate more than ±0.1 mm.

Dot Model
Check an arbitrary channel.

Recorder Settings

Set the input to the 2 V range, and −2.000 V to 2.000 V.

Procedures

Input values equivalent to 0, 25, 50, 75, and 100%, and record.
Measure the recorded results using a Glass scale (or equivalent), then confirm that
the results meet the criteria below.
Criteria

Input

Criteria

−2 V

0 ±0.3 mm

−1 V

25 ±0.3 mm

0 V

50 ±0.3 mm

1 V

75 ±0.3 mm

2 V

100 ±0.3 mm

Note

These criteria do not include the measurement accuracy. When taking the measurement
accuracy into account, the criteria are measurement accuracy ±0.3 mm.

Z-Fold Chart Paper Feed Accuracy

Procedures

1.

Turn ON the power switch while pressing the

MENU

and

CH UP

keys. Menu = A/D

Adjust is displayed.

2.

Select Menu = Test, and press the

key.

3.

Choose Test = Print and press the

key.

4.

Select Menu = Chart, and press the

key.

The recorder starts feeding paper, then stops after approximately 500 mm.
Time ticks are printed between the start and stop of paper feeding. Check that
the space between the two time ticks (d1 + d2 + d3 in the figure below) satisfies
the criterion below. Measure using a Glass scale (or equivalent).
Criterion: 500 ±0.4 mm

Time ticks

10 mm

d2 (Number of scale

intervals X 10 mm)

d3 (measured with

a Glass scale)

d1 (measured with

a Glass scale)

* Select using the key. You can hold down the SHIFT key while pressing key to reverse the
key’s action.

3.4 Test Procedures