Dead band (pen model), Pen offset (pen model) – Yokogawa µR20000 User Manual

Page 28

Advertising
background image

3-15

SM 04P02B01-01E

Testing

3

Dead Band (Pen Model)

Procedures

Check the accuracy of all pens. Use standard pens (or equivalent).

*

A “standard pen” is one whose tip does not fluctuate more than ±0.1 mm.

1.

Turn ON the power switch while pressing the

MENU

and

CH UP

keys. Menu =

A/D Adjust is displayed.

2.

Select Menu = Test, and press the

key.

3.

Choose Test = Dead Band and press the

key.

The dead band measurement pattern is automatically recorded near 0%, 50%,
and 100%. Measure the following portions of the recorded results using a Glass
scale (or equivalent), then confirm that the results meet the criteria.
Criterion: 0.36 mm or less

d

* Select using the key. You can hold down the SHIFT key while pressing key to reverse the
key’s action.

Pen Offset (Pen Model)

Procedures

Check all pens. Use standard pens (or equivalent).

*

A “standard pen” is one whose tip does not fluctuate more than ±0.1 mm.

1.

Turn ON the power switch while pressing the

MENU

and

CH UP

keys. Menu =

A/D Adjust is displayed.

2.

Select Menu = Test, and press the

key.

3.

Choose Test = Dead Band and press the

key.

The dead band measurement pattern is automatically recorded near 0%, 50%,
and 100%.
Measure the following portions of the recorded results using a Glass scale (or
equivalent), then confirm that the results meet the criteria.

Criteria

Item

Criteria

A

2.1 ±1 mm

B

4.2 ±1 mm

C

6.3 ±1 mm

A B C

* Select using the key. You can hold down the SHIFT key while pressing key to reverse the
key’s action.

3.4 Test Procedures