3 adjustments to the dot model, Adjusting the a/d converters – Yokogawa µR20000 User Manual

Page 38

Advertising
background image

4-7

SM 04P02B01-01E

Adjustments

4

4.3

Adjustments to the Dot Model

Adjusting the A/D Converters

Overview

Enter the value determined for each range, perform measurement, then write the
adjusted values to the A/D converter.

Instruments Used

Instrument

Specifications

DC voltage generator

Accuracy: ±0.005% of setting + 1 µV or better

Resistors

10 Ω, 50 Ω, 100 Ω, 150 Ω, 300 Ω, Accuracy: 0.01% or better

Channel

Enter the Zero value for channel 1 and the Full value for channel 2, and perform the
adjustment.

Procedure

Connection

UUA

When DC range

DC Voltage

standard

DECADE

RESISTANCE

BOX

Channel 1

H

L

UUA

When RTD or Cu range

Channel 1

b

DC Voltage

standard

DECADE

RESISTANCE

BOX

Channel 2

H

L

Channel 2

+

+

B A

b B A

(Zero value)

(Zero value)

(Full value)

(Full value)

UUA: Unit Under Adjustment

Input Value
Same as the pen model. See page 4-2.