6 zero-point calibration – LumaSense Technologies 3434i SF6 Leak Detector User Manual
Page 151

Chapter 12
______________________________________________________________________
_____________________________________________________________________________
BE6028-14
3434i SF6 Leak Detector
LumaSense Technologies A/S
Page 151 of 191
Fig.12.9 Illustration of the calibration data stored in the filter banks
of filters “A” to “E” after a zero-point calibration task
12.6
Zero-point Calibration
To calculate the concentration offset factor for a filter (during a
zero-point calibration) it is necessary to have a humidity gain
factor for the filter, and vice versa. If a humidity gain factor is not
stored in each of the filter banks, which are active during a zero-
point calibration, then it is necessary to perform a combined zero-
point and humidity-interference calibration task. The set-up of this
.
If it is only a newly-installed optical filter which needs to be
calibrated, then the same procedure used for an individual filter must
be used, see
. In this way you can protect your original
concentration offset factors.