LumaSense Technologies 3434i SF6 Leak Detector User Manual
Page 48

Chapter 4
______________________________________________________________________
_____________________________________________________________________________
BE6028-14
3434i SF6 Leak Detector
LumaSense Technologies A/S
Page 48 of 191
COMPENSATION FOR WATER VAP. INTERFERENCE
MONITORING TASK STARTED 2011-01-16 09:50
MEASUREMENT CHAMBER TEMPERATURE 29.6°C
SAMPLE INTEGRATION TIME W: 5s
A: 5s B:___s C:___s D:___s E:___s
A:Sulphur hexaflouride B:________ C:___________
D: ___________ E:__________ W: WATER VAP
A: 7.42E+00 B: ________ C:___________
D: ____________ E:
________ W: 8.13+00
The last display above shows the up-dated overview of the gas
concentrations measured during the most recent measurement cycle.
This display does not appear, after the monitoring task has been
stopped.
Data collected during a monitoring task is stored in the Display
Memory of the Monitor and therefore, before a new monitoring task
is started, all data stored in this memory is deleted. Display Memory
has finite size, and therefore there is a limit to how much data can
be stored in it. Once this memory has been filled-up all subsequent
data collected during the task will start to overwrite the data already
collected during the start of the monitoring task. Only in this event
will the date displayed after the text MONITORING TASK STARTED
on the above screen be different from the earliest date on the
display (shown below) which give the most up-to-date overview of
the monitoring task:
DATA IN DISPLAY MEMORY RECORDED FROM
2014-01-16 09:50 TO 2011-01-16 10:04
4. Press u.
This causes the following text to appear on the screen:
WATER VAPOUR MEASURED
Use q to scroll through the following screen displays:
MONITORING PERIOD IS NOT PRE-SET