running there is a positive flow of gas out of the flow meter.
This will ensure that the calibration gas is not diluted by
atmospheric air.
The following text appears on the screen:
RESULTS NOT YET AVAILABLE
PLEASE WAIT
Once the first measurement result is available the screen text
illustrated above is replaced by a screen displaying the
measurement results – for example the following:
W : µ : 5.34 µV σ : 212 nV 3 40.4 °C
PRESS ENTER WHEN RESULTS ARE STABLE
As soon as measurement results are available, they not only
appear on the display (see
) but are automatically
printed out on the printer, if the Monitor has been set up to print
. Remember to turn off the data log
when the calibration is complete.
Each time a sample of gas is drawn into the analysis cell the
signal is measured using both optical filter “A” and the water-
vapour filter (“W”) but, due to lack of space on the display, the
Monitor only displays the signal measured using the water-vapour
filter. The data-log print-out, however, shows the signal measured
using both the water-vapour filter and filter “A”.
A maximum of 6 measurements are stored in the Monitor during
any calibration. The number (n) indicates how many
measurements are stored. When 7 measurements have been
performed the number (n) will show the number 6 because the
very first measurement is overwritten by the 7
th
measurement so
that only 6 measurements are still stored. When 8 measurements
have been completed the number (n) will also show 6 because the
results of the first and second measurements have been
overwritten by the 7
th
and 8
th
measurement results …and so on.
2. Let the Monitor continue measuring the span gas A until the
temperature is stable. Look at the print-put of the average and
standard deviation measurements for (1) the water-vapour and
(2) the filter “A”. When these values have stabilised for both
filters, span calibration measurements do not need to continue.
3. Press
¿
.
This causes the following text to appear on the screen: