103

Oscilloscope Modules Specifications and
Characteristics

16534A Characteristics

Horizontal (Time)

Time base ranges

0.5 ns/div to 5 s/div

Time base resolution

10 ps

Delay range
• pretrigger

• -32 K x sample period

• posttrigger

• 320 ms or 1.6E7 x sample period, whichever is greater

Time interval measurement accuracy

±{(0.005% of

∆ T) + (2E–6 x delay setting)

forsampling rates other than maximum,

+ [0.15/(sample rate)]}

for bandwidth-limited signals [signal
rise time > 1.4/(sampling rate)], on a
single card, on a single acquisition

Time interval measurement accuracy

± [(0.005% of

∆T) + (2E–6 x delay setting) + 300 ps

for 2, 3, or 4 Agilent 16533As or 16534As
operating on a single time base, for
measurements made between channels
on different cards, at maximum
sampling rate

Trigger

Trigger level range (See notes)

±1.5 x full scale from center of screen

Trigger modes
• Immediate

• Triggers immediately after arming condition is met

• Edge

• Triggers on rising or falling edge on channel 1 or

channel 2

• Pattern

• Triggers on entering or exiting a specified pattern

across both channels

• Auto condition

• Self-triggers if trigger is not satisfied within

approximately 50 ms after arming

• Events delay

• The trigger can be set to occur on the nth occurrence

of an edge or pattern, n

≤ 32000

• Intermodule

• Arms another measurement module or activates the

port out BNC connector when the trigger condition is
met

Notes:
•

Specifications apply only within ± 10° C of the temperature at which the most recent calibration was performed.

•

Specifications apply only after operational accuracy calibration is performed in the frame in which the
oscilloscope module is installed.

•

Display magnification is used below 56 mV full scale. For sensitivities from 16 mV to 56 mV full scale, full scale is
defined as 56 mV.