110

Pattern Generation Modules Specifications
and Characteristics

Clock Cable Characteristics Without a Clock Pod

The Agilent 16720A and 16522A clock cables without a clock pod provide an ECL terminated
(1 K

Ω to –5.2V) differential signal (from a type 10E164 driver). These are usable when received by a

differential receiver, preferably with a 100

Ω termination across the lines. These signals should not be

used single ended due to the slow fall time and shifted voltage threshold (they are not ECL compatible).

100

Ω

Clock In

10E116

7

8

100

Ω

Wait 1, 2, 3 IN

10H125

11, 13, 15

12, 14, 16

–3.25 V

215

Ω

215

Ω

Clock Out

10E164

–3.25 V