6. Cleaning

Clean the seat and the housing as required using a domestic cleaning agent or commer-
cially-available disinfectant. Follow the manufacturer’s instructions.
Under no circumstances use abrasive or acid cleaners, white spirit, benzene or the like for
cleaning. Such substances can damage the high-quality surfaces.

7. What do I do if…

… no weight display comes on?
– Is the scale switched on?
– Check the batteries.

0.0

does not appear before weighing?

– Switch off the scale with the Start key and start the scale again – there must not be

any load on the scale – and only its feet should be in contact with the floor.

… one segment is illuminated constantly or not at all?
– The relevant segment has a fault. Inform Servicing.

… the display shows

?

– Battery voltage is running low. Change the batteries in the next few days.

bAtt

appears in the display?

– Batteries are empty. Put in new batteries.

SToP

appears in the display?

– Maximum load has been exceeded. Take the weight off the scale.

… the display flashes?
– If you have just activated the Hold or Tare function, please wait until the scale has de-

tected a steady weight value. The display will then stop flashing.

– If there has been no activation of a function beforehand, remove the load from the scale

and wait until

0.0

is displayed, then weigh again.

… the display

Er

and a number appear (e.g.

Er:12

)?

– Switch off the scale with the Start key and start the scale again. The scale will then

work normally again.
If this is not the case, disconnect the power supply by briefly removing the batteries. If
this measure is equally unsuccessful, inform the Service department.