Setup menu – DeFelsko PosiTector 6000 v.6.2 for Basic (1) and Standard (2) models User Manual

Page 11

Advertising
background image

11

-zeros all on-screen

Statistics and HiLo tabulations.

is above the

Hi limit. The LED will blink red if

readings are outside limits. Press (+) to clear

HiLo

readings.

Reset restores factory settings and returns
the Gage to a known, out-of-the-box
condition. It is handy when settings have been
changed, if the Gage behaves unusually, or if a
calibration adjustment is not possible for some
reason. The following occurs:

- all stored measurements are erased.

- calibration adjustments are cleared and returned

to the Gage’s factory calibration settings.

- This symbol appears on the display: It

disappears if a calibration adjustment is made by
the user.

- menu setting are returned to the following:

Hi Res = OFF Memory = OFF
Cal Lock = OFF Statistics = OFF
Hi Lo Alarm = OFF N Lock = OFF (FN models only)

A more thorough

Reset can be performed when the

Gage is powered down by holding the (+) button
until the Reset symbol appears. This is handy
when the Gage fails to power-up or operate properly.
It performs the same function as a menu Reset with
addition of

Units = microns, Flip Display = Normal and

Language = English.

-Keep the Gage away from metal during a

Reset.

Setup Menu

Setup Menu

NOTE: