Three steps ensure best accuracy…

1.Calibration - typically done by the manufacturer

or a qualified lab

2.Verification of Accuracy - as done by the user

3.Adjustment - to a known thickness

Calibration

Calibration is the controlled and documented
process of measuring traceable calibration
standards and verifying that the results are within the
stated accuracy of the Gage. Calibrations are
typically performed by the Gage manufacturer or by
a certified calibration laboratory in a controlled
environment using a documented process.