8

thicker value. These values should be on either side
of the thickness range to be measured.

1.Select

2 Pt Adjust from the Cal Settings menu.

2.Press (+) to select the number of readings to be

used to obtain an average on the thinner item,
typically 3 to 10 readings. The greater the variation
between readings, the more readings should be
taken to obtain an average.

3.Repeatedly measure the thinner item. The Gage

will wait for 2 seconds on the surface to allow the
user to correctly position the probe on the surface.
After the last measurement the Gage will calculate
and display a thickness value which represents the
average of all the readings taken using the factory
calibration settings.

4.Lift the probe from the surface and adjust the

displayed reading down (-) or up (+) to the known
thickness value of the thin item. Press to accept
this value.

5.Repeat steps 2 - 4 for the thicker item.

When selected, the icon will appear and the
current calibration settings are “locked” to prevent
further user adjustments.